| Antoni Nargol, M.D. in U.K. discusses Metal Ions and ASR XL and ASR revisions |
|
|
|
|
August 28, 2011 Tony Nargol: "We are seeing patients with no pain and severe osteolysis of pelvis with ASR’s. The critical investigation is metal ions testing. In UK it is exploding and each centre that had not seen the problem is steadily seeing failures. We are at 100% failure for ASR XL 44 and 46 cups and 50% for ASR XL.
Below is Failure numbers of ASR and ASR XL by Cup Inclinations
******WARNING***** EXTREMELY GRAPHIC SURGERY PHOTOS BELOW, DO NOT SCROLL DOWN IF YOU DO NOT WANT TO SEE LIVE SURGERY PHOTOS*** "The MHRA guidance was if metal ions are >7 ppb then surgeons should scan and investigate. However the detail is missing : normal is 2ppb , so 7 was way above normal and this means you are likely to be in trouble. However this number was only for resurfacing hips when the bearing surface was wearing. The problem with Total hip replacements like ASR XL is usually the bearing is fine , but it comes off the taper. Usually cobalt in preference to chromiuim and hence we prefer to observe cobalt. Many patients do not get to even a level of 7ppb before pain and severe damage sets in. It is difficult to get a big level of ions from a small taper.
40% of our ASR XL are between 4 and 7 and 66% of our Corail Pinnacle
36mm MoM hips between 3.1 and 7 ppb of cobalt. Our highest ion case from another centre had 230 ppb cobalt and chromium
Photo above is of a patient who had an ASR XL with pseudotumour , patient had mild pain and was still working , he had had the implant in him for 5 years, he came from Ireland to have his revision. Patient with a Corail Pinnacle metal on metal same problem
I then asked him about the fact that there is a wear in period where our levels WILL be higher the first year or two that we have our implants, here is his response. Tony Nargol: "Yes there is a run in for a year , but it is worth checking at a year and then repeat at 18 months if concerned. ions should not be above 4 at 18 months . However we are great believers in serial testing to see the trend." I then asked him also WHEN did he start seeing the failures happening, was it longer out and maybe that is why the U.S. surgeons are not yet seeing the negative results that he has seen because most of them have only been implanting the ASR for around 4 years, even Dr. Bose started implanting ASR's only like 4 1/2 years ago, and here is his response.
Tony Nargol: "Yes the ASR fails mostly after 4 yrs then
accelerates after that.
A few more very alarming photos of the ASR taper stem
He is seeing a LOT more failures in ASR XL's which are THR's with
a tapered stem. So please note that the numbers below are different,
in one area he is talking about ppb with BLOOD results, in another area
he is referring to ppb in HIP Aspirations. Mr. Tony Nargol recommends revision for patients with the following. Vicky : whom do we revise1, Pain : on Harris hip score mild or worse. 2, high metal ions cobalt >4 ppb 3, a positive scan These three raise our concern to move to next level of investigation. Even if only 2 out of three we then perform a hip aspiration. Hip aspiration gives a clue to the problem. we are presenting our results soon to show a key predictor of trouble is volume of fluid on aspiration. COLOUR OF ASPIRATE : IF white or yellow or mucky grey : needs a revision irrespective of all else if combined with big volume >20mls volume of aspirate : if red and volume not great like <10 mls one can wait but keep close eye and may repeat, but ions super high then revise. Hip metal ion levels : they should be around 50ppb , but if ten times then a problem. (THIS IS NOT BLOOD LEVELS) So anyone with rising ions >20ppb we are recommending revising. This decision i have taken only in last couple of weeks.
I will not revise if around 4 or 5 etc if in no pain.
40% of our ASR XL are between 4 and 7 and 66% of our Corail Pinnacle
36mm MoM hips between 3.1 and 7 ppb of cobalt. Tony" And to make this fair, I have asked some other surgeons that also are very experienced Hip surgeons that implanted ASR's to give their responses to the above data. Here they are, more will be added as I receive them.
On Aug 31, 2011, at 6:54 PM, Vijay Bose wrote: As I have indicated earlier I have revised only one ASR so far out of the 300 +. This was also due to cup slippage and not MoM problems. 4 other patients have pain / discomfort. Two of them have normal ions and most probably have ALVAL and thus not related to ASR.
The other two have marginally elevated levels but normal on imaging. They have refused revision so far.
I am at a loss to explain this disconnect as I know Tony personally and he is an outstanding surgeon.
On Sep 20, 2011, at 11:53 AM, John Antoniou wrote: |